  Type MZ30 - Reagent unit  
 
     
 
  Type Description  
  The reagent unit supplies the flow injection analysis within a Type 8905 system with the correct reagents. Up to three reagents are stored in a bottle holder for this. The device checks whether the correct reagent bottles are inserted. This function is ensured by the integrated barcode reader. The barcode provides additional information on the expiry date. The operator receives a message if a reagent is faulty. The level is detected through the weight of each bottle. In this case, the operator also receives a message if the reagents need to be topped up. The reagent unit can be installed in a system housing or attached to the mounting plate in an analysis cabinet. If an MZ30 without electronics is used, there is no barcode detection. The level is calculated based on the pump function. * Only applies to devices with barcode reader and weight sensors (product with electronics).  
     
  Features  
  « Module for storing and monitoring the reagents for the fully-automatic operation of the flow injection analysis modules  
  « Records the reagent level and indicates a required replacement in due time  
  « Checks the correct reagent type and storage life  
  « Fully EDIP compatible for easy system integration  
     
 
 
 
  Downloads  
  Data Sheet